High-Level Soak Disinfection
High-Level Disinfection is required for devices that contact non-intact skin.
In order for cleaner to be effective, all visible residue must be removed during the cleaning process. Follow the cleaning instructions as given in Manual cleaning Instructions to clean the probe before performing disinfection.
Do not soak probes in cleaner for longer than stated by the chemical manufacturers instructions for use. Extended soaking may cause probe damage and early failure of the enclosure, resulting in possible electric shock hazard.
- Prepare a basin with the high-level disinfectant per the manufacturer’s instructions for use. Be sure to follow all precautions for storage, use and disposal. Observe specifically soak times and dilution rates.
- Immerse cleaned and dried probe in the disinfectant for the time specified by the chemical manufacturer and ensure no air bubbles are trapped.
- Thoroughly rinse the probe to remove traces of the disinfectant. Follow the disinfectant manufacturers instructions.
- Thoroughly dry all surfaces of the device using a sterile, lint-free wipe or cloth, changing wipes/ cloths when necessary to ensure the device is completely dry.
Note
High-level disinfectants may be available as powder products. Follow the manufacturer’s instructions for use for preparing the disinfecting solution from the powder product.