Release Notes
2.11.0
New Features
Multi-Language Support
- Multi-language support with French, German, and Spanish language options
- Language selection available on login screen and main menu
- All help descriptions now display in your selected language
- Date pickers show months and days in your selected language
- Reports now display in your selected language (PDF and Excel formats)
Generator Integration Enhancements
- Generator status display in the System Status bar showing current generator state
- Improved exposure control for Sedecal and iRay generators during image acquisition
- Technique adjustment options during generator calibration
- Progress display showing calibration status during iRay calibration
- Generator Settings Dialog closes automatically after saving changes
Imaging and Anatomy Features
- New PHARYNX anatomy option for equine imaging
- Updated Carpus anatomy settings for improved image processing
Patient Management Improvements
- Automatic filling of Owner Last Name when adding patients from worklist
User Experience Enhancements
- Better handling when image files are missing with clear "Image Not Found" message
- Improved calibration workflow with helpful guidance messages
- Clear warning message when importing settings that requires system restart
Bug Fixes
- Fixed an issue where Summit generator performance would degrade after system reset
- Fixed an issue where configurable behaviors would not apply properly when adding patients from worklist
- Fixed an issue where Required Fields settings were not being properly enforced in patient dialogs
- Fixed an issue where image enhancement settings would not work properly in Tuning Workbench
- Fixed an issue where backup restore would fail between different system versions
- Fixed an issue where backup space calculation was incorrect when restoring images
- Fixed an issue where technician names would not appear correctly in billing reports
- Fixed an issue where angle measurement text would not stay in place after moving it
- Fixed an issue where dropdown background color was incorrect in light mode theme
- Fixed an issue where restart time dropdown showed incorrect 15-minute intervals
2.10.2
Bug Fixes
Fixed an issue where the Add Patient dialog settings were not applied correctly when adding a patient from the worklist.
2.10.1
Bug Fixes
- Fixed a bug that caused systems integrated with Summit generators to become unresponsive to user input under specific conditions.
2.10.0
New Features
Workflow & DICOM Improvements
- Added a configuration to automatically set the Referring Physician for a study from the Worklist - reducing manual entry.
- Introduced new dialogs for adding and editing DICOM Servers (Worklist, Storage, MPPS), designed to ensure reliable and consistent updates.
- Automatically includes an additional Patient History DICOM tag from the Worklist when creating studies - improving exported DICOM data completeness.
Patient Management
- Users can now specify additional required fields in the Add/Edit Patient dialogs - providing greater control over required data collection.
Imaging Enhancements
- Improved SoundVu Fusion tuning parameters for Third Carpal Bone views.
Performance & Stability
- The application now creates a diagnostic crash report (minidump) when the application crashes to aid in faster troubleshooting.
- The application now automatically restarts daily for optimal performance. The time of day is configurable in the Advanced Options (default is 2 AM).
- Added additional error handling when loading images from disk to reduce load failures and improve stability.
Library & Integration Updates
-
Updated MVVMDialogs to version 9.1.2.
-
Updated SedecalSerial library to version 3.1.1.146.
-
Replaced NPOI with ClosedXML for more reliable Excel file handling.
Bug Fixes
- Fixed an issue where the wrong technique was set on Summit generators when powered on from the acquire screen.
- Fixed a display issue with the DNS radio button selection text.
- Fixed an issue where the annotation toolbar could not be shown even when annotations were enabled.
- Fixed a bug where entering an incorrect Wi-Fi password appeared to connect successfully, but the connection was not actually established.
- Fixed an issue where Sedecal generators could not fire after a power cycle.
- Fixed an exception that was logged when adding a measurement annotation.
- Fixed an issue in the Add User Dialog where the Save button appeared active initially but could not be used.
- Fixed an issue in demo mode where switching between real and demo detectors did not update the demo acquire button visibility.
- Fixed a Fusion detector issue where entering and then exiting the acquire review screen before readiness could leave the detector in an invalid state.
- Fixed an issue where acquisition did not respect the configured order of shots in the shot list.
- Fixed a display issue in the Add/Edit Patient dialogs on systems with screen scaling other than 100%.
- Fixed a potential memory leak when exporting images to a DICOM server.
- Fixed a bug where Batch Exports could conflict if multiple batch jobs were started quickly in succession.
- Fixed a possible memory leak in the DICOM Queue.
- Fixed a Windows handle leak when rendering annotations and overlays during image export.
2.9.0
New Features
- Added support for integrating the iRay VDR632 generator.
- Users can now filter MWL searches by Issuer of Patient ID.
- Users can now manage network adapters from within the application.
- Users can now connect to Wi-Fi networks from within the application.
- Users can now pair Bluetooth LE devices from within the application.
Bug Fixes
- Fixed an issue that caused extra messages to be sent to the Summit generator.
- Fixed an issue that caused HD and HD+ detectors to sometimes fail to initialize correctly.
- Fixed an issue that caused the application to crash when trying to reprocess images captured before the 2.5.0 update.
- Fixed an issue that caused the application to crash when trying to crop images that contain annotations.
- Fixed an issue that caused the incorrect version to be set in the database when running older installation scripts.
- Fixed an issue where manual restoration continued restoring images after database restoration failed.
2.8.0
New Features
- Study Comparison - Users can now load and compare a secondary study side-by-side with a primary study in the Acquire Review screen, with independent navigation and overlay controls for each.
- User management tasks are now processed individually, ensuring that each action (like adding, updating, or deleting a user) completes successfully before moving to the next. This helps prevent partial changes and keeps your data consistent.
- Support for new Fusion 1717 v2, Fusion 1417 v2, and Fusion 1114 v2 detectors.
Bug Fixes
- Resolved an issue where password were visible when adding / editing users.
- Resolved an issues where the Sedecal Generator integration option was not available.
2.7.0
New Features
- Sedecal Integration
- Users can now manage passwords.
- DICOM Nodes can now support host names as well as IP address.
- Add grams (g) and ounces (oz) as options for Default Weight Units.
- The Emergency Patient Workflow can now be disabled.
- The Image Tuning Workflow can now be disabled.
- When adding a Study, if the patient ID entered already exists, the user can add a study to the existing patient.
- New Use Exact Birthday configuration option, when enabled, requires the user to enter an exact date instead of using the Years and Months dropdown.
- Shotlists now have "Handles" for easier reordering.
- Incoming Species names from MWL servers can now be mapped.
- Acq Profiles automatically selects the Default Species if one is specified.
- Worklist and MPPS servers can now be exported/imported.
Bug Fixes
- Fixed an issue where Worklist Items with Scheduled Procedure Step Status of "Completed" or "Canceled" were displayed.
- Fixed an issue that could cause a crash if the user clicked the Learn More button on an update notification when not logged in and tried to access other areas of the application.
- Fixed an issue where Patient and Owner Names could be cut off in the Patient tile.
- Fixed an issue in the Reject Report (Excel) where Patient Name and Acquisition data was swapped.
- Fixed an issue in the Adv. Billing Report (Excel) where the file would be exported to the incorrect directory.
- Fixed an issue in the AAHA Report (Excel) where the Ma data was not being exported.
- Fixed an issue where the Search Filters had a very small hit box when using touch.
- Fixed an issue where Breed and Sex were not being populated in the AIS Consultation.
- Fixed an issue where Sex may not populate on Worklist items.
- Fixed an issues where HD and HD+ detectors were unable to acquire images immediately after calibration.
2.6.0
New Features
- Added Cobb Angle Annotation.
- Support for HD 1717 and HD+ 1717 detectors.
- Support for SD 1717 detectors.
- Added a new Advanced Configuration option Patient Search Result Limit. This limits the number of search results returned from a local search.
- When the Default Species is set to equine, the calibration instructions have been customized.
- Added a notification when the detector's batter reaches a low charge.
- Added a notification when the detector's batter reaches a critically low charge.
- Added a new optional Additional Details field to the Reject Image dialog. This is also shown on the Reject Report.
- Add new 'Keeneland Racing' and 'Keeneland Non-Racing' Protocols. The shots in these protocols have the correct description for Keeneland. These protocols replace the old 'Keeneland Repository' protocol.
Bug Fixes
- Resolved an issue where a detector can go to sleep during the calibration process, causing the dialog to be unable to close normally.
- Resolved an issue where errors during calibration were not displayed to the user and only showed as canceled.
- Resolved an issue where the Shotlist is out of order after reordering certain shots.
- Resolved an issue where the user was not notified that their Image Processing license was in Grace.
- Resolved an issue where the DICOM Tag (0028,0301) BurnedInAnnotations always had the Value "NO", it now contains "YES" if overlays are burned into the image on export.
- Resolved an issue where the Reject Reason drop down would change sizes depending on the selection.
- Resolved an issue where Acquisition could fail to start when the detector's battery was critically low.
- Fixed an issue where two dialogs could be shown simultaneously for the same error when starting acquisition.
- Resolved an issue where creating a backup including images could fail due to high disk space usage. Backups including images will no longer zip up all image files and will export them directly to the selected destination.
2.5.1
Bug Fixes
- Resolved an where the software would crash when trying to send a DICOM N-CREATE when the patient already exists on the system.
2.5.0
New Features
- Calibration no longer requires the user to Trigger the detector. Detectors will automatically detect exposure during calibration.
- Simplified the Self Test to show the user what self tests have been run, what is currently running and the results of each.
- New Diagnostic tool to clear detector storage i.e. clear untransferred images from detector.
Bug Fixes
- Resolved an issue where the Configuration Option to Include Worklist Constraints in Query was being ignored.
- Resolved an issue where Detector Battery level was not being updated as often as expected.
- Resolved an issue where the RSSI was not being updated in a timely manner for wireless detectors.
- Resolved an issue where Study Date and Study Time Overlays would not display in the acquisition screen.
- Resolved an uncommon issue where Acquisition could be enabled while in the Create Shotlist View or Add/Edit Shots views.
- Resolved an issue where the Shotlist would not scroll at top or bottom when dragging a shot.
- Resolved an issue where the Patient's name would not update on Create Shotlist View when edited.
- Resolved an issue where detectors could exhaust internal storage space.
- Resolved an issue where Image processing license could be reported as expired when it was not.
2.4.0
Bug Fixes
- Resolved an issue that prevented studies from being deleted by the Watermark system if they contained any rejected images.
- Resolved an issue in the Network service when setting wifi profiles where when the panel profile is already set up but more profiles exist, it will remove all profiles instead of just the additional profiles.
2.3
New Features
- Updated Large Animal Overlay Landscape Preview Image
- Log Version at startup
- Limited AE title fields to 16 characters to conform to the DICOM standard.
- Added System and Application event logs to the Data Collector.
- Added SQL Error logs to Data Collector.
- Added Descriptions for Fusion Errors to logs.
- DICOM Library updated to 5.0.3
Bug Fixes
- Resolved an issue that caused COMM_ATTACK errors with Summit generators when Wattage of requested technique exceeded generator limits.
- Resolved an issue where DICOM Sends got stuck when the server dropped connection at a critical point.
- Resolved an issue where the Generator Error dialog would not close.
- Resolved an issue where series level overlays did not display for newly acquired images.
- Resolved an issue where Image Laterality overlay would not update when the View Position was changed.
- Resolved an issue where DICOM Sends failed when the Calling AE was longer than 16 characters.
- Resolved an issue where clearing Annotations would not apply to the database.
- Resolved an issue where image data was loaded spuriously when exiting the Acquire Review screen.