Skip to main content

Worklist

Worklist

In order to search for Worklist Items, you will need to configure a Worklist Server. Only one worklist server is allowed.

Add a Server

Click the Add Button. Enter a name for the server. Enter the IP address of the Server. Enter the AE Title of the Server. Enter the port number of the server. Enter in the Scheduled Station AE Title if your workflow or Worklist server require one.

By default SmartDR only queries Worklist servers for DX modality. You can include CR Modalities by selecting the Supports CR checkbox. To query all modalities, select the Supports any modality.

danger

You can only have one Worklist Server configured.

Press the Save button to save the Server.

Delete a Server

To delete a server, select the server you wish to delete and press the Delete Button.

Ping

You can Ping the IP address to verify IP Connectivity with the Server By pressing the Ping button. The results of the Ping will be displayed in a popup. Press the Close button to dismiss the popup.

Ping Results

Echo

You can verify DICOM connectivity by pressing the Echo Button. The results will be displayed in a popup. Press the Close button to dismiss the popup.

Echo Result

Include Worklist Constraints in Query

If this option is set to True, constraints derived from the search settings will be used to query the worklist server. The default value is False.

Worklist Query Result Limit

This is the number of results that will be displayed when a worklist search is performed. By default it is set to 100. Options are:

  • 50
  • 100
  • 250
  • 500
  • 1000
  • 2500
  • 5000

Only Show Scheduled Worklist Patients

If this option is set to True the results will only contain studies that have not been started.

Enable application-Side Result Filtering

This option determines if constraints such as the Patient ID filter and settings such as the Supported Modalities value are included in worklist queries sent to the worklist device. If True is selected, then a query does not include constraints and the worklist device returns all patient records. With this method, results are filtered from the SmartDR application side rather than the query side. If False is selected, then a query includes constraints and the worklist device returns only patient records that meet the constraint criteria. Default: False