HD/HD+

Name
This is used to identify the Detector to the users.
IP Address
This is the IP address of the detector. By default this is set to 192.168.100.8.
Interface
This is the interface the system will communicate with the detector. By default it is set to Wireless.
info
If there are issues registering detectors, double check this value. HD/HD+ detectors cannot switch between Wired and Wireless modes.
Correction Directory
This is the directory that contains the defect, gain, and offset files for the detector. This will be supplied by a Sound Tech Support or Installer.
Register
Press this button to register the detector. This button will be disabled until a Correction Directory has been selected.
note
The detector must be connected and powered on.